Specs Digital Infrared Temperature Sensor FT series

Sensor head

Model

FT-H10

FT-H20

FT-H30

FT-H50

FT-H40K

FT-H50K

Image

Type

Mid to low temperature Small-spot

Mid to low temperature Mid-range

Mid to low temperature Long-range

Mid to low temperature Ultra long-range

High temperature Mid-range

High temperature Ultra long-range

Rated temperature range

0 to 500°C*1

0 to 1350°C*1

Displayable temperature range

-50 to +520°C

-50 to +1400°C

Detecting distance

Unlimited *2

Measuring distance /
View diameter (example)

17.5/ø6.3 mm
35/ø1.5 mm
52.5/ø8 mm

60/ø8.5 mm
120/ø6 mm
180/ø14.5 mm

200/ø15 mm
400/ø18 mm
600/ø33 mm

500/ø18 mm
1500/ø30 mm
3000/ø75 mm

100/ø7.5 mm
150/ø6 mm
300/ø11.5 mm

500/ø18 mm
1500/ø30 mm
3000/ø75 mm

Sight

2-point visible laser beam*3

Detecting element

Thermopile

Detecting wavelength

8 to 14 µm

Repeatability

±0.5°C

±3°C

Emissivity (ε) correction

0.10 to 1.99 (0.01 step)

Environmental resistance

Ambient temperature

-10 to +55 °C (No freezing)

Relative humidity

35 to 85 % RH (No condensation)

Vibration resistance

10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z axis

Shock resistance

500 m/s2, 10 times in each direction of each of X, Y and Z axis, 60 times in total

Material

Case: Reinforced glass plastic, Infrared collector lens: Germanium, Laser transmitter: Polyarylate
Lens tube*4: Aluminum, Cable: Vinyl chloride, Mounting bracket: SUS304, Mounting screw: Stainless steel

Material

Mounting bracket

SUS304

Accessories

Mounting bracket

Weight

Approx. 120 g

Approx. 150 g

*1 Repeatability is guaranteed within the rated temperature range.
*2 Place at a distance so that object appears 1.5 times larger than the view diameter.
*3 Visible semiconductor laser beam wavelength 655 nm, Class 1 Laser Product [IEC60825-1, FDA (CDRH) Part1040.10]. The laser classification for FDA (CDRH) is implemented based on IEC 60825-1
in accordance with the requirements of Laser Notice.
*4 No lens tube on models FT-H10, FT-H20, and FT-H30.

Back to top

Model

FT-50A

FT-50AW

FT-55A

FT-55AW

Image

Type

DIN-rail mounting type

Panel mounting type

Rating

Power voltage

12 to 24 V DC ±10% ripple (P-P) 10% or less

Power consumption

Normal

1400 mW (For 12 V), 1600 mW (For 24 V)

Powerconsumption

Eco mode

1150 mW (For 12 V), 1350 mW (For 24 V)

Display power

4 + 1/two-digit 7-segment LED, dual-display (red/green) display
(character height in the upper display (red): 8 mm, character height in the lower display (green): 5.7 mm), Display updating cycle: 10/s

Operation indicator

Red LED x 3 (control output1/control output2/timing input indicator)

Display resolution

0.1°C or 1°C (when using the FT-H10/H20/H30), 1°C (when using the FT-H50/H40K/H50K)

0.1°C or 1°C (0.2°F or 1.8°F) (when using H10/H20/H30), 1°C (1.8°F) (when using H50/H40K/H50K)

0.1°C or 1°C (when using the FT-H10/H20/H30), 1°C (when using the FT-H50/H40K/H50K)

0.1°C or 1°C (0.2°F or 1.8°F) (when using H10/H20/H30), 1°C (1.8°F) (when using H50/H40K/H50K)

Hysteresis

Variable

Response

HSP, 30, 100, 200, 500, 1000, or 5000 ms can be selected (In HSP: 10ms typ., 15 ms max.)

Bank selection

Switch from the timing input line and analog output line to bank input, up to 4 banks can be set

Pink and Purple wires can be configured for Bank Inputs #1,2 respectively for up to 4 total banks.

Switch from the timing input line and analog output line to bank input, up to 4 banks can be set

Pink and Purple wires can be configured for Bank Inputs #1,2 respectively for up to 4 total banks.

Bank specifications

For each bank, emissivity (during temperature mode) and the set value × 2 can be set

Emissivity (valid in temperature mode) and limit settings x 2 can be set for each bank.

For each bank, emissivity (during temperature mode) and the set value × 2 can be set

Emissivity (valid in temperature mode) and limit settings x 2 can be set for each bank.

External input

Timing input

Input time: 2 ms min.

Bank input

Input time: 20 ms min.

Control output

NPN open collector x 2 channels (N.O./N.C. selectable), 100 mA max. (40 VDC max.), residual voltage: 1 V max.

Analogue output

4 to 20 mA, maximum load resistance: 260 Ω The upper- and lower-limit values of the analog output range can be set optionally.

Environmental resistance

Ambient temperature

-10 to +50 °C (No freezing)*1

Relative humidity

35 to 85 % RH (No condensation)

Vibration resistance

10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z axis

Material

Main unit, Cover: Polycarbonate
Keycaps: Polyester elastomer
Cable: Vinyl chloride
Mounting bracket: SPCC

Main unit, Cover: Polycarbonate
Keycaps: Polyester elastomer
Cable: Vinyl chloride
Panel mounting bracket: Polyacetal
front protection cover: Polycarbonate

Accessories

DIN fixing bracket, power cable, unit cover sticker

DIN anchoring fixture, power cable, unit cover sticker

Panel mounting bracket, protection cover, power cable, unit cover sticker

Weight

Approx. 85 g

*1 Ambient temperature when using the analog output is -10 to 45 °C

Back to top

Model

FT-50AP

FT-50AWP

FT-55AP

FT-55AWP

Image

Type

DIN-rail mounting type

Panel mounting type

Rating

Power voltage

12 to 24 V DC ±10% ripple (P-P) 10% or less

Power consumption

Normal

1400 mW (For 12 V), 1600 mW (For 24 V)

Powerconsumption

Eco mode

1150 mW (For 12 V), 1350 mW (For 24 V)

Display power

4 + 1/two-digit 7-segment LED, dual-display (red/green) display
(character height in the upper display (red): 8 mm, character height in the lower display (green): 5.7 mm), Display updating cycle: 10/s

Operation indicator

Red LED x 3 (control output1/control output2/timing input indicator)

Display resolution

0.1°C or 1°C (when using the FT-H10/H20/H30), 1°C (when using the FT-H50/H40K/H50K)

0.1°C or 1°C (0.2°F or 1.8°F) (when using H10/H20/H30), 1°C (1.8°F) (when using H50/H40K/H50K)

0.1°C or 1°C (when using the FT-H10/H20/H30), 1°C (when using the FT-H50/H40K/H50K)

0.1°C or 1°C (0.2°F or 1.8°F) (when using H10/H20/H30), 1°C (1.8°F) (when using H50/H40K/H50K)

Hysteresis

Variable

Response

HSP, 30, 100, 200, 500, 1000, or 5000 ms can be selected (In HSP: 10ms typ., 15 ms max.)

Bank selection

Switch from the timing input line and analog output line to bank input, up to 4 banks can be set

Pink and Purple wires can be configured for Bank Inputs #1,2 respectively for up to 4 total banks.

Switch from the timing input line and analog output line to bank input, up to 4 banks can be set

Pink and Purple wires can be configured for Bank Inputs #1,2 respectively for up to 4 total banks.

Bank specifications

For each bank, emissivity (during temperature mode) and the set value × 2 can be set

Emissivity (valid in temperature mode) and limit settings x 2 can be set for each bank.

For each bank, emissivity (during temperature mode) and the set value × 2 can be set

Emissivity (valid in temperature mode) and limit settings x 2 can be set for each bank.

External input

Timing input

Input time: 2 ms min.

Bank input

Input time: 20 ms min.

Control output

PNP open collector x 2 channels (N.O./N.C. selectable), 50 mA max. (30 VDC max.), residual voltage: 1 V max.

Analogue output

4 to 20 mA, maximum load resistance: 260 Ω The upper- and lower-limit values of the analog output range can be set optionally.

Environmental resistance

Ambient temperature

-10 to +50 °C (No freezing)*1

Relative humidity

35 to 85 % RH (No condensation)

Vibration resistance

10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z axis

Material

Main unit, Cover: Polycarbonate
Keycaps: Polyester elastomer
Cable: Vinyl chloride
Mounting bracket: SPCC

Main unit, Cover: Polycarbonate
Keycaps: Polyester elastomer
Cable: Vinyl chloride
Panel mounting bracket: Polyacetal
front protection cover: Polycarbonate

Accessories

DIN fixing bracket, power cable, unit cover sticker

DIN anchoring fixture, power cable, unit cover sticker

Panel mounting bracket, protection cover, power cable, unit cover sticker

Weight

Approx. 85 g

*1 Ambient temperature when using the analog output is -10 to 45 °C

Back to top

Model

FT-S1*1

Image

Air supply

Recommended flow rate

15ℓ/min

Withstanding pressure

1 MPa or less (at inlet)

Conforming tube diameter

ø6 mm

Material

Housing: zinc, cap: aluminum,
cable packing and O-ring: NBR,
fitting: Brass nickel-plated

Weight

Approx. 700 g

*1 FT-S1 does not provide cooling.

Back to top

Model

FT-S2

Image

Infrared
transmissivity

85 % or more*1

Material

Cap: aluminum,
lens: germanium,
O-ring: NBR

Weight

Approx. 32 g

*1 Infrared transmissivity for wavelength of 8 to 14 µm.

Back to top

Model

OP-91147

Image

Emissivity

0.95

Allowable temperature limit

180°C (356°F)

Size

Width: 50 mm , length: 10 m

Weight

Approx. 145 g

Back to top